Model Wise Description
PTX-3000-U - Process Transmitter with universal input (8 user selectable)
PTX-3000-F - Process Transmitter with single fixed input
PTX-3000-T - Process Transmitter with two same input
Technical Specification
Input Sensor
J,K,R,S,B thermocouple or RTD (Pt-100) or mV, mA signal (Input type not listed here can also be provided)
Range
Full range of the selected input desired measuring range can be chosen from range selection table..
Accuracy
Software linearized to +/-1 oC for thermocouple input and +/-0.1oC for Pt-100 input.
For Voltage and Current the accuracy is absolute to the input.
Power Supply
24VDC +/- 10% @ 150mA
Current Output 1
4-20mA proportional to the specified range of the user selected input. (12 bit resolution). 2 wire
Current Output 2
4-20mA proportional to the specified range of the second specified input. (12 bit resolution). For PTX-3000-T
Digital Serial Output
Isolated 2 wire RS485 on MODBUS RTU protocol in SLAVE mode
Isolation
4-20mA and RS485 outputs are isolated from DC Power supply.
Isolation voltage > 1.5KV DC.
Loop Resistance
RLmax = 400 Ohms - for each output.
Display
4 digit - 0.3 red seven segment LED .
4 digit each for PTX-3000-T
Connection
2 or 3 wire for sensor input, 2 wire for DC supply and
2 x 4-20mA for current outputs, 2 wire for RS-485 outputs.
Cable Entry
Through suitable size gland for supply and input/output.
Enclosure (Size)
ABS plastic IP-65 (120 x 80 x 55) mm. Download PDF Catalog Click Here
|